Use the following procedure to remove and replace the measurement side if
necessary.
Equipment required:
Phillips drive, #2
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2–1).
2. Unplug the fan cable J18
Figure 7–9. Unplugging the Fan Cable
3. Unplug DMC cable
(Figure
7–9).
(Figure
7–10).
Servicing

Measurement Side Removal

48iQ Instruction Manual
7-9