Servicing
DMC Optical Bench
7-44
48iQ Instruction Manual
5. Using a 7/64 hex drive, unfasten two #6-32 socket cap head screws.
6. Pull preamp board assembly outwards.
Figure 7–47. Preamp/Detector Removal
7. Replace preamp/detector as needed and assemble in reverse order.
Note When reinstalling preamp board assembly, make sure the o-ring is
seated properly.