Download Print this page

Ndir Bench - Thermo Scientific 48iQ Instruction Manual

Carbon monoxide analyzer

Advertisement

NDIR Bench

Thermo Scientific
The NDIR Bench screen provides status and alarms related to the bench
module. If an item being monitored goes outside the lower or higher alarm
limit, an alarm is activated.
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>NDIR
Bench
The NDIR Bench screen contains the following information:
Across:
NDIR Bench: This column lists items associated with the NDIR
bench module.
Value: Displays the current value for each item.
Low Alarm: Displays low alarm status for each item.
High Alarm: Displays high alarm status for each item.
Units: Displays units for each item.
Down:
S/R: Displays the current sample/reference reading.
Sample Intensity: Displays the current sample intensity reading.
Reference Intensity: Displays the current reference intensity reading.
Bench Pressure: Displays the current bench pressure reading.
Flow: Displays the current sample flow reading.
Instrument Temperature: Displays the current instrument
temperature reading.
Operation
Settings
48iQ Instruction Manual
3-47

Advertisement

loading

This manual is also suitable for:

117080-00