Figure 7-38. Motor And Filter Optical Switch Replacement (Pt 3) - Thermo Scientific 48iQ Instruction Manual

Carbon monoxide analyzer

Advertisement

Thermo Scientific
Note Replace the optical switch if needed. You don't have to remove the
optical switch to replace the motor or filter.
5. Using #1 Phillips drive, unfasten two #4-40 flat head screws to remove
the optical cover.
6. Using #1 Phillips drive, unfasten one #2-56 pan head screw to remove
the optical switch.
Figure 7–38. Motor and Filter Optical Switch Replacement (pt 3)
DMC Optical Bench
48iQ Instruction Manual
Servicing
7-33