Calibration
3-30
48iQ Instruction Manual
The Point 1 screen allows the user to view and set the selected calibration
Point 12
point. The point 2 screen functions the same way. Therefore, the following
example of point 1 screen applies to point 2 as well.
Home Screen>Calibration>Advanced Calibration>2-Point O
Calibration>Continue to 2-Point O
Calibration Point 1
The 2-Point O
information:
Edit O
Current O
Edit O
button, the Point 1 Concentration value is updated.
Calibrate O
Note Ensure that the sensor has been sampling the gas for at least 1 minute
before pressing the Calibrate O
Calibration Point 1 screen contains the following
2
Span Concentration: User enters O
2
Concentration: Read only. After selecting the value on the
2
Concentration button and after pressing the Calibrate O
2
Point 1: Calibrates point 1 coefficient.
2
Point 1 button.
2
Calibration>2-Point O
2
span concentration.
2
2
2
2