Download Print this page

Multipoint Calibration - Thermo Scientific 48iQ Instruction Manual

Carbon monoxide analyzer

Advertisement

Calibration
Calibration
Multipoint
Calibration
4-6
48iQ Instruction Manual
The Multipoint Calibration should be performed initially when the
instrument is received. The user can choose to calibrate with one, two, or
three points. It is suggested to perform a 3-point Multipoint Calibration.
The Multipoint Calibration is divided into three calibration points. The
range is user definable and the following calibration points are
recommended for the user selected range:
Point 1: 80% of range
Point 2: 50% of range
Point 3: 20% or range
Note If performing a multipoint calibration after a "bad" calibration or
changing ranges, it is recommended to start off with default values for the
Cal point coefficients.
To set default values, from the Home Screen select Calibration>Advanced
Calibration>Multipoint Calibration. From Multipoint Calibration, select
Reset Points to Default Values. The instrument will automatically reset to
default values.
Thermo Scientific

Advertisement

loading

This manual is also suitable for:

117080-00