Carbon monoxide analyzer

Advertisement

Operation
Settings
3-48
48iQ Instruction Manual
Bench Temperature: Displays the current bench temperature
reading. User can adjust low and high alarm limits.
Bench Temperature too Low: Displays OK/Fail for bench
temperature too low.
Bench Temperature too High: Displays OK/Fail for bench
temperature too high.
Motor Speed too Low: Displays OK/Fail for motor speed too low.
Motor Speed too High: Displays OK/Fail for motor speed too high.
IR Source Current too Low: Displays OK/Fail for IR source current
too low.
IR Source Current too High: Displays OK/Fail for IR source current
too high.
IR Detector Bias too Low: Displays OK/Fail for IR detector bias too
low.
IR Detector Bias too High: Displays OK/Fail for IR detector bias too
high.
Bench Thermistor Open: Displays OK/Fail for bench thermistor
open.
Bench Thermistor Short: Displays OK/Fail for bench thermistor
short.
Module Thermistor Open: Displays OK/Fail for module thermistor
open.
Module Thermistor Short: Displays OK/Fail for module thermistor
short.
Board Communication: Displays OK/Fail for board communication
status.
Power Supply: Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
3.3V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
5.0V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
15V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
24V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.