Servicing
DMC Optical Bench
7-40
48iQ Instruction Manual
Figure 7–43. Motor Installation Shaft Alignment
6. Let instrument sample zero air for about 90 minutes.
7. From the Home screen, choose Settings>Measurement
Settings>Advanced Measurement Settings>Optical Bench
Settings>Continue to Initial S/R Calibration.
8. Follow steps in Chapter 4, "Calibration" for
"Initial
S/R".
Thermo Scientific