Optical Bench Removal; Figure 7-35. Dmc Removal From Instrument - Thermo Scientific 48iQ Instruction Manual

Carbon monoxide analyzer

Advertisement

Servicing
DMC Optical Bench
Optical Bench
7-30
48iQ Instruction Manual
Use the following procedure to remove the optical bench from the
instrument case.
Removal
Equipment required:
Phillips drive, #2
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2. Unplug DMC cable J6 (bottom).
3. Disconnect plumbing as shown.
4. Using a #2 Phillips drive, unfasten six captive hardware.
Figure 7–35. DMC Removal from Instrument
2–1).
Thermo Scientific

Advertisement

loading