Troubleshooting
Troubleshooting Guide
6-2
48iQ Instruction Manual
Problem
Solenoid current out of range
(option)
Pump current out of range
No output signal (or very low
output)
Calibration drift
Non-linear response
Excessive response time
Alarm – Instrument
Temperature
Alarm – Bench Temperature
Possible Cause
Action
Micro SD Card
If Micro SD card was just replaced,
Programming
re-install the old one. If the problem
is fixed, request a replacement
Micro SD card.
Sticking or damaged
Reset solenoid via Settings>Health
solenoid
Check>Status and Alarms>Valve and
Pump Resets screen. If damaged,
replace solenoid valve block.
Damaged or dirty
Reset pump via Settings>Health
pump
Check>Status and Alarms>Valve and
Pump Resets. Inspect and refurbish
pump. If pump motor is damaged,
replace pump.
No sample gas
Check input sample flow.
reaching the analyzer
Ruptured pump
Rebuild pump head.
diaphragm
Blocked sample
Unplug power cord. Clean or replace
capillary
capillary.
Line voltage
Check to see if line voltage is within
fluctuations
specifications.
Defective pump
Rebuild pump.
Clogged capillaries
Unplug power cord. Clean or replace
capillary.
Clogged sample air
Replace filter element.
filter
Incorrect calibration
Verify accuracy of multipoint
source
calibration source gas.
Leak in sample probe
Check for variable dilution.
line
Partially blocked
Unplug power cord. Clean or replace
sample capillary
capillary.
Hang up/blockage in
Change element.
sample filter
Low line voltage
Check to see if line voltage is within
specifications.
Fan failure
Replace fan if not operating
properly.
Dirty fan filter
Clean or replace filter.
Overheating PCB
Locate defective PCB reporting the
error and replace if needed.
Defective bench
Check 10K ohm thermistor and