Oxygen Sensor Capillary Replacement; Figure 7-60. Replacing The Oxygen Sensor Capillary - Thermo Scientific 48iQ Instruction Manual

Carbon monoxide analyzer

Advertisement

Servicing
Optional DMC Oxygen Sensor
Oxygen Sensor
Replacement
7-56
48iQ Instruction Manual
Use the following to replace the oxygen sensor capillary.
Capillary
Equipment required:
Ball-end hex drive, 3/32
1. Remove the oxygen sensor DMC from the instrument. Refer to
"Oxygen Sensor
2. Using a 3/32 ball-end hex drive (angle alignment of screws), unfasten
two #4-40 cap screws.
3. Unlock tube clamp.
4. Unplug tube after unlocking tube clamp.
Figure 7–60. Replacing the Oxygen Sensor Capillary
Removal" on page 7-51.