Advanced
Calibration
Thermo Scientific
The Advanced Calibration screen provides several additional ways to
calibrate the instrument and view the calibration history. See Chapter 4
"Calibration" for further instructions on how to run a calibration.
The following screens show the advanced screens in single range mode and
dual or auto range mode. (The only difference between the screens, are the
words "High" and "Low".) For more information about range modes, see
"Range Mode
Selection" on page 3-76.
Home Screen>Calibration>Advanced Calibration (single range mode
with O
option)
2
Home Screen>Calibration>Advanced Calibration (dual or auto range
mode with O
option)
2
The Advanced Calibration screens contain the following information:
Manual Calibration: The user manually adjusts the background or span
coefficient.
Multipoint Calibration: Up to three gas concentrations (cal-points) may
be calibrated. Three cal-points will give the most accurate readings over
the entire range.
Operation
Calibration
48iQ Instruction Manual
3-17