Figure 7-45. Heater Removal - Thermo Scientific 48iQ Instruction Manual

Carbon monoxide analyzer

Advertisement

Servicing
DMC Optical Bench
7-42
48iQ Instruction Manual
3. Turn bench over, backside of optical bench.
4. Using #1 Phillips drive, unfasten five #4-40 pan head screws.
Figure 7–45. Heater Removal
5. When replacing with new heatsink or thermistor, heatsink compound
is required. Apply a thin and even coat of heatsink compound
underneath.