Startup

Use the following procedure when starting the instrument.
1. Turn the power ON.
2. Allow 90–120 minutes for the instrument to stabilize. During the time
that the instrument is warming up, the mode "warm up" is displayed
on the gas mode button in the title bar, and the concentration
calculation is turned off. To disable warm up, go to
Settings>Configuration.
3. Set instrument parameters such as operating ranges and averaging times
to their appropriate settings. For more information about instrument
parameters, see the "Operation" chapter.
4. Before beginning the actual monitoring, perform a multipoint
calibration as described in the "Calibration" chapter.
Figure 2–8. Front Panel and Touchscreen Display
Installation and Setup
Startup
48iQ Instruction Manual