Thermo Scientific 48iQ Instruction Manual page 111

Carbon monoxide analyzer

Advertisement

Thermo Scientific
Current Sample Int. (Hz): Displays the current sample intensity in
Hz.
Current Reference Int. (Hz): Displays the current reference intensity
in Hz.
Down:
Manual Entry: Shows current values. If detector gain cell is pressed,
user can manually set the detector gain value.
Auto Cal: When Start is pressed, the auto-calibration process is
initiated. Please allow up to 5 minutes for calibration to complete.
User can stop calibration by pressing the Stop button.
End Cal: When Stop is pressed, the auto calibration is interrupted
and the value does not change.
Default Values: When pressed, the default gain value is saved.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue to Initial S/R Calibration
The Initial S/R Calibration screen contains the following information:
Across:
Description: Defines the actions the user can do.
Initial S/R.: Displays the initial S/R value.
Current S/R.: Displays the current S/R value.
Down:
Manual Entry: Shows current value. If the initial S/R cell is pressed,
user can manually set the initial S/R value. The instrument should
be sampling zero gas when changing the S/R.
Operation
Settings
48iQ Instruction Manual
3-83