Filter Wheel
Replacement
Thermo Scientific
Use the following procedure to replace the filter wheel.
Equipment required:
Hex drive, 5/32 and 5/64
Phillips drive, #2
1. Remove optical bench. Refer to
30.
2. Remove purge wheel housing. Refer to
Removal" on page 7-34.
3. Using 5/32 hex drive, unfasten two #10-32 socket cap head screws.
4. Using #2 Phillips drive, unfasten #8-32 pan head screw to gain access
to #8-32 set screw.
5. Using 5/64 hex drive, loosen #8-32 set screw two or three turns only.
6. Carefully slide chopper wheel off motor shaft.
7. Replace filter wheel and align motor and filter wheel. Refer to
and Filter Wheel
Alignment" on page 7-39.
"Optical Bench
Removal" on page
"Purge Wheel Housing
48iQ Instruction Manual
Servicing
DMC Optical Bench
7-
"Motor
7-35