Troubleshooting
Troubleshooting Guide
6-4
48iQ Instruction Manual
Problem
Alarm – 5V/24V Step Board
Alarm – Conc.
Alarm – Analog I/O
Alarm – Digital I/O
Alarm – Auto Bkg Cal/Check
Alarm – Auto Span
Cal/Check
Alarm – Bench Temp
Thermistor Open
Alarm – Bench Temp
Thermistor Short
Alarm – Motor Speed too
low
Alarm – Motor Speed too
high
Alarm – Module Thermistor
Open
Alarm – Module Thermistor
Short
Alarm – Bench Temp
Thermistor Open
Possible Cause
Action
Defective DMC PCB
Replace DMC board.
Cable connection
Check the cable connections to that
Step POL board.
Concentration has
Check to ensure range corresponds
exceeded range limit
with expected value. If not, select
proper range.
Concentration low
Check user-defined low set point. Be
sure the minimum trigger is set as
desired.
Defective PCB
Replace Analog board.
Defective PCB
Replace Digital board.
Incorrect high alarm
Verify the high limit is correct via
limit
Settings>Status and
Alarms>Concentrations screen.
Instrument
Recalibrate the instrument.
background
calibration failed
Incorrect high alarm
Verify the high limit is correct via
limit
Settings>Status and
Alarms>Concentrations screen.
Instrument span
Recalibrate the instrument.
calibration failed
Cable connection
Check connection from heater to
DMC board. Reseat if needed.
Cable connection
Check connection from heater to
DMC board. Reseat if needed.
Cable connection
Check connection from motor to
DMC board. Reseat if needed.
Defective Heater
Replace motor as needed.
Defective DMC board
Replace board as needed.
Cable connection
Check connection from motor to
DMC board. Reseat if needed.
Defective Heater
Replace motor as needed.
Defective DMC board
Replace board as needed.
Defective DMC board
Replace board as needed.
Defective DMC board
Replace board as needed.
Cable connection
Check the cable connection from the
thermistor to the DMC board.