Carbon monoxide analyzer

Advertisement

Operation
Settings
Optical Bench Settings
3-82
48iQ Instruction Manual
The Optical Bench Settings screen allows the user to manually or
automatically set the detector gain and initial S/R (sample/reference) value.
The S/R value should be calibrated prior to other instrument calibrations.
For more information, see Chapter 4, "Calibration".
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue to Detector Gain
Calibration
The Detector Gain Calibration screen contains the following information:
Across:
Description: Defines the actions the user can do.
Detector Gain: User can manually set or automatically calibrate the
detector gain.