What If...?
•
Nothing happens when you follow the operating instructions:
-
Check the system power.
-
Check for loose cable connections.
•
The AWS goes back to monitor OK prompt:
-
Resetting or disconnecting/reconnecting the bar code scanner may cause the workstation to
return to the monitor OK prompt. In this case:
Type the command go <Enter> to return to the normal working mode.
•
The scanner does not read a bar code symbol:
-
Be sure the scanning system is programmed to read the type of bar code you are trying to scan.
Refer to section
-
Be sure the bar code symbol is not defaced.
-
Be sure that you are holding the scanner correctly to scan the symbol, as illustrated below. The
scan head must touch and cover every bar and space on the symbol. You can tilt the scanner for-
ward or backward. Practice will quickly show you how to hold the scanner.
9-worklist.fm
3-2 Set default configuration on page
Right
Senographe DS Acquisition System
92.
Page no. 93
Operator Manual 5307907-3-S-1EN
Wrong
Worklist