Error messages
Message on
Gantry Readout
X12 Tube head control
failure
X13 X-ray beam adjust-
ment error
X14 X-ray beam form
calibration error
X15 Collimator com-
mand error
X16 Collimator control
failure
X17 Collimator soft-
ware error
X18 Filter command
error
Chapter 16
Explanation
The tube focal spot size cannot be
set:
- The tube head device has failed.
- It is not possible to perform an
image acquisition.
The tube head device that posi-
tions the tube to get the correct X-
ray beam position has detected
an error.
- The tube focal spot size cannot
be set correctly.
- It is not possible to perform an
image acquisition.
The tube focal spot size cannot be
set correctly.
- It is not possible to perform an
image acquisition except during
calibration.
The collimator device that posi-
tions the collimator blades
detected an error.
- The Field Of View cannot be set
correctly.
- It is not possible to perform an
image acquisition.
The collimator device has failed:
- The Field Of View cannot be set
correctly.
- It is not possible to perform an
image acquisition.
An error occurred on the collima-
tor device CPU:
- The Field Of View and the filter
cannot be set correctly.
- It is not possible to perform an
image acquisition.
The Filter device that positions
the filter detected an error.
- The filter cannot be set correctly.
- It is not possible to perform an
image acquisition.
Page no. 196
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1) If you hear the tube tilt motor, the system is try-
ing to recover itself.
2) If the message "S61 X-ray beam not ready" is
present for more than 60 seconds on the x-ray
control console, power OFF/ON from the X-ray
console.
3) If the problem persists, shut down the system
from the Browser.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) If you hear the tube tilt motor, the system is try-
ing to recover itself.
2) If the message "S60 X-ray beam quality error" is
present for more than 10s on the X-ray control con-
sole, power OFF/ON from the X-ray console.
3) If the problem persists, shut down the system
from the AWS Browser.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
Note ALL messages and contact GEMS Field Ser-
vice.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.