GE Senographe DS Operator's Manual page 188

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Error messages
Message on
Gantry Readout
H08 Power supply unit
comm. Failure
H09 Reset the system.
Reload the SW
H10 Reset the system.
Reload the SW
H11 Reset the system.
Reload the SW
H12 Reset the system.
Reload the SW
H13 Rotation communi-
cation failure
Chapter 16
Explanation
The system is set in emergency
mode:
- The Power unit is not communi-
cating; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- An error occurred on the Gantry
CPU; the system is set in "halt"
state. (occurred only during soft-
ware upgrade)
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- An error occurred on the Gantry
CPU; the system is set in "halt"
state. (occurred only during soft-
ware upgrade)
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- An error occurred on the Gantry
CPU; the system is set in "halt"
state. (occurred only during soft-
ware upgrade)
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- An error occurred on the Gantry
CPU; the system is set in "halt"
state. (occurred only during soft-
ware upgrade)
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- The Rotation device is not com-
municating, the system is set in
"halt" state.
- It is not possible to perform any
movement or image acquisition.
Page no. 188
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
16-errors.fm

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents