GE Senographe DS Operator's Manual page 168

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Error messages
Message on
X-ray Console
E26 AWS/SENO comm.
failure
E35 SYSTEM SHUT-
DOWN STARTED
E36 CART PWR FAIL-
URE - SHUTDOWN
I05 Exposure aborted by
mAs level
I06 Abort: Detect sig.
out range
I12 Maximum mAs = xxx Maximum available mAs are
I13 IDC was not ready
Chapter 16
Explanation
AWS system cannot communi-
cate with Generator.
A complete system power OFF
has been started.
The Generator will power off sup-
plies to the Detector, Gantry and
Generator in a few minutes.
A system shutdown has been
started because an UPS error
has been detected or communi-
cation with UPS has been lost.
Exposure aborted because not
enough mAs available to con-
tinue exposure.
Exposure aborted because
brightness received by detector is
not in range to continue expo-
sure.
equal to xxx.
Image Detection Control system
was not ready to perform an
image acquisition.
Page no. 168
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1. Power OFF/ON from the X-ray Console.
2. If the problem occurs again at the next acquisi-
tion, power OFF the whole system and wait for the
end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
Stop system usage and wait for the end of the
shutdown sequence.
1. Stop system usage and wait for the end of the
shutdown sequence.
2. Check the mains power supply.
3. Check (on UPS panel) that the batteries are
charging. Do not attempt to restart the system until
the batteries have recharged.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
1. Press the Exposure Interrupt button.
2. Reposition the patient.
3. Start exam again.
4. If the problem continues, note ALL messages
and contact GEMS Field Service.
5. Nevertheless, you can continue to use the sys-
tem with manual acquisition.
1. Press the Exposure Interrupt button.
2. Reposition the patient.
3. Start acquisition again.
4. If the problem continues, note ALL messages
and contact GEMS Field Service.
5. Nevertheless, you can continue to use the sys-
tem with manual acquisition.
Information message only.
(If mAs exceeded, the exposure is aborted at 2 or 3
mAs if entire exposure cannot be made.)
1. Press the prep button again.
2. If the problem occurs again at the next acquisi-
tion, power OFF the whole system and wait for the
end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
16-errors.fm