GE Senographe DS Operator's Manual page 178

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Error messages
Table 2 Error messages displayed on the AWS monitor
Message on AWS screen
AA These messages relate to power on/off events:
AA1
Error: Could not connect to
the remote provider !
AA2
ATTENTION Do you really
want to SHUT DOWN the
complete system? This
action will stop the work-
station and the X genera-
tor!
AA3
End of remote connection
in progress
AA4
Image detection failure,
exam forbidden
AA5
Initialization not complete,
please wait
AA6
No Ups Power Backup
AA7
Senograph failure, exam
forbidden
AA8
Shutdown Process started
...
Chapter 16
Explanation
The specified remote host could
not be found. It may be tempo-
rarily off-line, its address may
have changed or been incor-
rectly entered, the application
on the remote provider may not
be running, or there may be a
network error.
This message appears as part
of the normal shutdown
sequence.
A GE Insite connection is being
closed.
An image detection subsystem
is not ready for acquisition after
start-up.
A system initialization process is
in progress.
A UPS error has been detected
or communication with the UPS
has been lost.
A Senographe (Gantry) subsys-
tem is not ready for acquisition
after start-up.
The AWS shutdown process
has been started because a
UPS primary power failure lon-
ger than 10 seconds has been
detected.
Page no. 178
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective action
1. Try again after a short wait.
2. If the message is repeated, check that
the server address is correct and that the
remote application is running.
3. If the problem persists, contact your
Field Service Representative.
Click Yes or No as required.
No action is required; you can now use
the system.
1. Wait to see if the problem clears.
2. If the problem persists, contact your
Field Service Representative.
Wait until system initialization is com-
plete.
1. Stop system usage and shut down the
system.
2. Check the mains power supply.
3. Check (on UPS panel) that the batter-
ies are charging. Do not attempt to
restart the system unless the batteries
have been recharged.
4. Restart the system.
5. If the problem persists, contact your
Field Service Representative.
1. Wait to see if the problem clears.
2. If the problem persists, contact your
Field Service Representative.
1. Wait for the end of the shutdown
sequence.
2. Check mains power supply.
3. Restart the system when power OK.
4. If the problem persists, contact your
Field Service Representative.
16-errors.fm