X-ray Console
21. Exposure button
Press and hold this button while the Exposure Enable lamp is lit to make an exposure. To stop the
exposure immediately, release the Exposure button.
A buzzer sounds at the end of exposure.
Note:
In AOP mode, and following preparation (anode rotation), exposure is made in two stages:
An absorption measurement exposure (the "pre-exposure"), to allow selection of the optimum kV,
focal track, and filter combination for the chosen priority. These parameters are then set automat-
ically. Maximum pre-exposure time is 70 ms (large focal spot) or 130 ms (small focal spot).
An exposure during which image acquisition is made.
The operator must maintain pressure on the Exposure button until the buzzer stops to
obtain proper exposure.
22. Compression release button
Press this button to release compression. The compression paddle automatically moves up to the
pre-programmed auto-decompression height.
Note:
If auto-decompression is selected in the medical set up menu, compression is released automati-
cally at the end of the exposure.