Ejecting a disk from the CD-R drive
As soon as the CD-R drive has been accessed for a save or restore operation, it is locked, and you can
no longer eject the disk by means of the eject button on the front of the drive.
You must first perform a Detach operation:
Click on the storage device icon in the Main Browser and select Detach in the drop-down
menu.
If the disk is not ejected automatically, press the eject button.
If Detach is displayed in grey and no save or restore operation is in progress, the archive
device is already unlocked. Press the eject button to open the drive.
If a fault occurs (for example trying to start a restore operation with a blank disk in the drive), the Detach
operation is performed automatically, and the disk is ejected.
! Notice:
NEVER press the eject button on the CD-R drive while a save or restore operation is in progress.