Error messages
Table 1 Error Messages displayed on the X-ray Console
Message on
X-ray Console
E01 Generator failure
E02 Focus bias failure
E03 Rotor failure
E04 Arm fatal failure
E05 Failure during
exposure
E06 Software error (gen
CPU)
E07 Please TURN OFF
the console !
E08 Power supply fail-
ure
E10 Heating failure
track 1
E11 Heating failure
track 2
E12 Checksum error
(gen)
E13 Gen./Arm commu-
nication fail
Chapter 16
Explanation
Problem reported in HV section
of generator.
Problem reported in tube focus
bias circuits.
Problem reported in anode
starter.
Generator detects error in dialog
with arm.
Current exposure aborted due to
failure in equipment.
An error occurred in the Genera-
tor CPU software.
A generator emergency failure
occurred but a second failure
prevented the system from shut-
ting itself down.
Supply of generator power cir-
cuits is not authorized.
Molybdenum focal track is unser-
viceable.
Rhodium focal track is unservice-
able.
Calibration parameters have
been modified.
Generator cannot communicate
with arm.
Page no. 166
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
Power OFF/ON from the X-ray Console. If the
problem continues, note ALL messages and con-
tact GEMS Field Service.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
inform GEMS Field Service.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
inform GEMS Field Service.
1) Note ALL messages.
2) Switch OFF the system from the X-ray console
to quickly switch off the generator.
3) Contact GEMS Field Service.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL
messages and contact GEMS Field Service.
You can still work with rhodium focal track.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
You can still work with molybdenum focal track.
Power OFF/ON from the X-ray Console. If the
problem continues, note ALL messages and con-
tact GEMS Field Service.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact
GEMS Field Service.