GE Senographe DS Operator's Manual page 172

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Error messages
Message on
X-ray Console
I50 Lock Jammed
I51 Lock Overheated
I90 Auto-Decompres-
sion is Off
S01 Aborted exposure
S02 Exposure aborted
by HV arcing
S05 Wrong selection of
kV/track couple
S07 Exposure aborted
by IDC
S08 Required power is
too high
Chapter 16
Explanation
The system is in degraded mode:
Gantry device configuration is not
a standard configuration. The
system functionality is not opti-
mum.
The system is in degraded mode.
The tube head fan has failed. As
a result, tube cooling is not effec-
tive and the tube thermal algo-
rithm could not prevent exposure
abort due to tube overheating.
The automatic
decompression mode has been
inhibited by the user.
Exposure in progress has been
aborted because the Exposure
Button was released by operator
before the end of exposure.
Exposure was aborted because
more than eight X-ray tube arcing
events occurred.
Selected kV/focal track is not
enabled (kV values are too low in
the Rh focal track).
The exposure was aborted by the
Image Detection Controller.
Power requested in Manual
Mode is too high.
Page no. 172
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1) If the installed paddle is not a Senograph DS
paddle:
A. Remove it.
B. Install a Senograph DS paddle.
2) If the installed paddle is a Senograph DS pad-
dle:
A. Remove it and reinstall it.
B. If the warning persists, check with another
Senograph DS paddle.
B1. If the warning disappears with another
Senograph DS paddle, the paddle that generated
the warning is not recognized by the system; the
thickness/force measurements with this paddle are
not accurate.
B2. If the warning persists with another Seno-
graph DS paddle, the thickness/force measure-
ment with all paddles are not accurate.
1) The system can continue to be used but please
reduce the cadence of use.
Add time between exposures and exams to
allow the tube to cool.
2) Contact GEMS Field Service to repair the tube
head fan.
If auto-decompress mode is wanted, switch modes
by changing the parameter in the Setup/Medical
menu. Otherwise no action is required; this infor-
mation is normally displayed.
Power OFF/ON from the X-ray Console.
Start exam again.
Power OFF/ON from the X-ray Console.
Start exam again.
Change the configuration.
1. Power OFF/ON from the X-ray Console.
2. If the problem occurs again at the next acquisi-
tion, power OFF the whole system and wait for the
end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
Reduce kV or mAs values.
16-errors.fm