GE Senographe DS Operator's Manual page 192

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Error messages
Message on
Gantry Readout
W06 Invalid language.
Default is used
W07 Control buttons
error
W08 A footswitch is
pressed
W09 Gantry Readout
failure
W10 Lift calibration
error
W11 Lift control error
W12 Lift end position
exeeded
W15 Rotation calibra-
tion error
W16 Angulation control
error
W17 Rotation control
error
Chapter 16
Explanation
The Gantry CPU received incor-
rect language information and will
use the default value.
The default language value is the
English language.
An error occurred on an arm con-
trol button keypad, which does not
affect the current system use.
An error occurred on a footswitch
device that does not affect the
current system use.
The gantry readout has failed.
The lift device is not well cali-
brated; no movement will be avail-
able until the lift device will be well
calibrated.
An error occurred during lift move-
ment that does not affect the
safety control but the speed pro-
file only.
The Lift movement has stopped at
a firmware limit, but should have
stopped before.
The rotation device is not cor-
rectly calibrated; no movement
will be available until the rotation
device has been correctly cali-
brated.
An error occurred on the Rotation/
Angulation clutch/brake control.
An error occurred during rotation
movement. It does not affect the
safety control, but only the speed
profile.
Page no. 192
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, continue to use the sys-
tem with English language set. Note ALL mes-
sages and contact GEMS Field Service.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
Contact GEMS Field Service.
Note ALL messages and contact GEMS Field Ser-
vice.
Note ALL messages and contact GEMS Field Ser-
vice.
1) Use lift slow speed in the upper and lower limit
area.
2) Note ALL messages and contact GEMS Field
Service.
Note ALL messages and contact GEMS Field Ser-
vice.
1) If an exam is currently in progress continue it,
using the same rotation incidence.
2) Move away from 0° and return to 0°; try again to
switch to rotation mode.
3) If the problem persists, move the compression
paddle up, switch to rotation mode, perform rota-
tion, and return to angulation mode.
4) If the problem persists, power OFF/ON from the
X-ray Console.
5) If the problem persists, note ALL messages and
contact GEMS Field Service.
Note ALL messages and contact GEMS Field Ser-
vice.
16-errors.fm