GE Senographe DS Operator's Manual page 77

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Quality Check
Aim of the Quality Check feature
The Quality Check feature requires the user to apply a qualification level to each image. This qualifica-
tion level is:
Used to trigger Auto Push and/or Auto Print
Used to generate image quality statistics over a given period of time (Repeat and Reject Analysis or
RRA)
How to configure Quality Check
Case 2: RRA disabled
Auto Push
Auto Print
Default image quality check
OK
NOT OK
Save
Initial decision: Do you need the RRA feature?
The RRA feature provides an alternative (and automated) method to the manual Repeat Analysis
method described in the Senographe DS QC Manual.
It is mainly intended for sites that follow MQSA regulations. For such sites, it is mandatory to perform a
repeat analysis at least quarterly.
It is also useful for other sites to generate and keep statistics on total repeat and reject rates, as well as
their respective reasons.
Case 1: RRA enabled
If you need the RRA feature, click Yes in the Enable Repeat and Reject Analysis module, then select the
cases for which you wish to make use of the Auto Print and Auto Push features.
Note:
Once RRA is enabled, the Default image quality check is set to OK and cannot be modified.
8-browser.fm
MedAppPrefs
Annotations
Auto Windowing
Enable Repeat and Reject Analysis
Page no. 77
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Case 1: RRA enabled
Quality Check
Auto Delete
Yes
Auto Push Repeated images?
No
Auto Push Rejected images?
Auto Print Repeated images?
Auto Print Rejected images?
Test Hosts
Names
Image Process.
Yes
No
Yes
No
Yes
No
Yes
No
Close
Browser
Chapter 8