GE Senographe DS Operator's Manual page 189

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Message on
Gantry Readout
H14 Rotation control
failure
H17 Stop motion con-
trol line failed
H18 Stop motion line
active
H19 Compression foot-
switch failure
H20 Lift footswitch fail-
ure
H21 Lift software error
H22 Halt. Uncontrolled
lift motion
16-errors.fm
Explanation
The system is set in emergency
mode:
- The Rotation device is failed; the
system is set in "halt" state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- Stop movement command fail-
ure; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- The stop movement button has
been pressed; the system is set in
"halt" state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- Compression footswitch failure;
the system is set in "halt" state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- Lift footswitch failure; the system
is set in "halt" state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- An error occurred on lift device
CPU; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- Uncontrolled lift movement; the
system is set in "halt" state.
-It is not possible to perform any
movement or image acquisition.
Page no. 189
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
Error messages
Chapter 16