Message on
X-ray Console
E14 Generator CPU/
INTG failure
E15 Console communi-
cation fail
E18 IDC/SENO Syn-
chro. fail
E19 IDC/SENO down,
DPS is OFF
E20 Check Main Distri-
bution Rack
E21 Detector supply not
supplied
E22 Conditioner not
supplied
E23 AWS cart not sup-
plied
E24 Failure before
exposure
E25 AWS/SENO Syn-
chro. fail
16-errors.fm
Explanation
A problem has been reported by
the Generator CPU.
Generator cannot communicate
with X-ray Console.
A problem occurred in synchroni-
zation between Image Detection
Controller system and Generator
during acquisition. The image
may be fully acquired, but dose
parameters will be missed (set to
0 in the image).
Next exposure is inhibited.
Image Detection Controller sys-
tem cannot communicate with
Generator.
The Detector Power Supply has
been powered OFF by the sys-
tem.
Supply of Control Station and
Detector environment circuits is
not authorized.
Supply of Detector circuits is not
authorized.
Supply of detector Conditioner
circuits is not authorized.
Supply of Control Station circuits
is not authorized.
Current image acquisition
aborted before exposure due to
failure in the equipment
A synchronization problem
occurred between AWS system
and Generator during acquisition.
Next exposure is inhibited.
Page no. 167
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
1. Power OFF/ON from the X-ray Console.
2. If the problem occurs again at the next acquisi-
tion, power OFF the whole system and wait for the
end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
1. Power OFF/ON from the X-ray Console.
2. If the problem occurs again at the next acquisi-
tion, power OFF the whole system and wait for the
end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
1. Power OFF/ON from the X-ray Console.
2. If the problem persists, note ALL messages and
inform GEMS Field Service.
1. Power OFF/ON from the X-ray Console.
2. If the problem persists, note ALL messages and
inform GEMS Field Service.
1. Power OFF/ON from the X-ray Console.
2. If the problem persists, note ALL messages and
inform GEMS Field Service.
1. Power OFF/ON from the X-ray Console.
2. If the problem persists, note ALL messages and
inform GEMS Field Service.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
1. Switch OFF the whole system and wait for the
end of the shutdown sequence.
2. Restart the system.
3. If the problem persists, note ALL messages and
inform GEMS Field Service.
Error messages