Message on AWS screen
AA9
Shutdown Process started
...
AA10
Shutdown Process started
...
AA11
System is down. System
will power off within 30
seconds: do NOT reboot
now! (if system does not
turn off, switch off UPS
manually)
AA12
System not ready, exam
forbidden
AA13
UPS battery needs replac-
ing. Please call service
AA14
UPS output overload.
Please call service
BB These messages relate to boot/reset events:
BB4
AWS reset in progress:
please wait
BB6
Communication with UPS
not established. Please
call service. System will
power off within 30 sec-
onds: do NOT log in now!
BB8
Remote connection in
progress : do not acquire
images
16-errors.fm
Explanation
The AWS shutdown process
has been started because com-
munication with the UPS was
lost for more than 30 seconds.
The AWS shutdown process
has been started because a
UPS low battery condition was
detected (battery power must be
available to guarantee safe
operation and shutdown in the
event of power failure).
The message appears at the
end of the normal shutdown
sequence, when the UPS is
about to shut down, to prevent
users from trying to reboot.
A system component is not yet
ready (after system start-up).
The UPS selftest has detected
that the battery requires
replacement.
An UPS overload condition has
been detected.
An AWS boot or reset sequence
is in progress.
The AWS may start an auto-
matic reset if it detects an inter-
nal acquisition device error.
This message appears at the
end of the power-up/boot
sequence, if communication
cannot be established with the
UPS.
A GE Insite connection is in
progress; no new images should
be acquired.
Page no. 179
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective action
1. Stop system usage and wait for the
end of the shutdown sequence.
2. Restart the system.
3. If the problem persists, contact your
Field Service Representative.
1. Stop system usage and wait for the
end of the shutdown sequence.
2. Check the mains power supply.
3. Check (on UPS panel) that the batter-
ies are charging. Do not attempt to
restart the system until the batteries
have recharged.
4. If the problem persists, contact your
Field Service Representative.
No action is normally required.
If the system fails to shut down com-
pletely (the AWS LCD monitor continues
to display the message after one or two
minutes), switch it off by pressing the
UPS OFF button at the front of the Gen-
erator Cabinet.
Wait until system initialization is com-
plete before requesting an acquisition.
Stop system usage and contact your
Field Service Representative.
Stop system usage and contact your
Field Service Representative.
Wait until system initialization is com-
plete.
If several automatic resets occur, contact
your Field Service Representative.
1. Wait for the end of the shutdown
sequence.
2. Try to restart the system.
3. If the problem is repeated, contact
your Field Service Representative.
No action is required; wait for completion
of the connection.
Error messages