GE Senographe DS Operator's Manual page 190

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Error messages
Message on
Gantry Readout
H23 Rotation brake fail-
ure
H24 Rotation software
error
H25 Halt. Rotation
Uncontrolled
H27 Compression
brake failure
H28 Control button fail-
ure on arm
H29 Control button fail-
ure on head
Chapter 16
Explanation
The system is set in emergency
mode:
- The arm rotation brake has
failed; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- An error occurred on the rotation
device CPU; the system is set in
"halt" state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- Uncontrolled Rotation move-
ment; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- The compression device is in
fault state; the system is set in
"halt" state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- An arm movement control button
on an arm keypad (behind the
image receptor, left or right) has
failed; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
The system is set in emergency
mode:
- An arm movement control button
on an X-ray tube head keypad
(left or right) has failed; the sys-
tem is set in "halt" state.
-It is not possible to perform any
movement or image acquisition.
Page no. 190
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Power OFF/ON from the X-ray Console.
2) Try to perform compression motion. If the prob-
lem persists, note ALL messages and contact
GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.