GE Senographe DS Operator's Manual page 183

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Message on
Gantry Readout
2D Large Localization
Paddle. Maximum rec-
ommended force is
20daN.
No defined paddle
Round Spot Magnifica-
tion paddle
Axillary paddle
19x23 paddle
Flexible 19x23 paddle
Round spot paddle
2D Spot Localization
Paddle. Maximum rec-
ommended force is
20daN
Square Magnification
paddle
16-errors.fm
Explanation
The system informs the operator
that the "Biopsy" paddle is present
and reminds the operator that it is
recommended to not exceed
20daN with this paddle.
The system informs the operator
that it does not detect a
Senographe DS paddle.
The system informs the operator
that the "Round Mag Stand" pad-
dle is present. This paddle should
be used with a magnification plat-
form.
The system informs the operator
that the "Axillary" paddle is pres-
ent.
The system informs the operator
that the "19x23 Standard" paddle
is present.
The system informs the operator
that the "19x23 Phlexis" paddle is
present.
The system informs the operator
that the "Round spot" paddle is
present.
The system informs the operator
that the "Spot Biopsy" paddle is
present and reminds the operator
that it is recommended to not
exceed 20daN with this paddle.
The system has detected that the
2D Spot localization paddle is
present and reminds the Operator
that it is recommended to not
exceed 20daN with this paddle.
Page no. 183
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
If the "Biopsy" paddle is not installed, an error has
occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
If you observe this message when a Senographe
DS paddle is present, contact GEMS Field Service.
If the "Round Mag Stand" paddle is not installed,
an error has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
If the "Axillary" paddle is not installed, an error has
occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
If the "19x23 Standard" paddle is not installed, an
error has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
If the "19x23 Phlexis" paddle is not installed, an
error has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
If the "Round spot" paddle is not installed, an error
has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
If the "Spot Biopsy" paddle is not installed, an error
has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
If the 2D Spot localization paddle is not installed,
an error has occurred:
1. Power OFF/ON from the X-ray Console.
2. If the problem persists, note ALL messages
and contact your Field Service Representative.
Error messages
Chapter 16