GE Senographe DS Operator's Manual page 184

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Error messages
Message on
Gantry Readout
Square Spot Paddle
19x23 Magnification
paddle
Unexpected paddle
type
Unknown paddle type
A02 Compression force
is applied
A03 Compression is not
operational
A04 Compression
status is unknown
Chapter 16
Explanation
The system informs the operator
that the "Square Spot" paddle is
present.
The system informs the operator
that the "Standard Mag Stand"
paddle is present. This paddle
should be used with a magnifica-
tion platform.
Paddle device error
The system informs the operator
that the installed paddle is not rec-
ognized.
Lift and rotation movements are
not allowed because a compres-
sion force is applied.
Lift, rotation and compression
movements are not allowed
because the compression device
is not operational. Compression
status is temporarily unknown.
A lift or rotation movement has
been requested while the Stereot-
actic Positioner is not ready.
Page no. 184
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
If the "Square Spot" paddle is not installed, an
error has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
If the "Standard Mag Stand" paddle is not installed,
an error has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
Note ALL messages and contact GEMS Field Ser-
vice.
If the paddle is a Senographe DS accessory:
1) Remove the paddle and insert it again.
2) If the message is still displayed after inserting:
A. Power OFF/ON from the X-ray console.
B. If the problem persists, use another
Senographe DS paddle.
C. If the problem persists, note ALL messages
and contact GEMS Field Service.
Decompress the breast (move the paddle holder
up) if you want to perform a rotation or a lift move-
ment.
None
1) If the Stereotactic Positioner has just been
installed, wait for the "READY" message on the SP
display before pressing a footswitch or control
movement button.
2) If "COMMUNICATION NOT ESTABLISHED" is
displayed on the SP display, perform the correc-
tive action for this error. When the system has
recovered, move the gantry as required.
3) Check the X-ray Console for error messages
and perform appropriate corrective actions.
4) If the problem persists, power OFF/ON from the
X-ray Console.
5) If the problem persists, note ALL messages and
contact GEMS Field Service.