GE Senographe DS Operator's Manual page 169

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Message on
X-ray Console
I14 AWS was not ready
I15 Console locked
I16 Collimator locked
I17 AWS/SENO comm.
lost
I18 Collimator/Console
locked
I19 IDC/SENO comm.
lost
I21 Heating failure track
1
I22 Heating failure track
2
16-errors.fm
Explanation
AWS system was not ready to
perform an image acquisition.
The X-ray Console is locked due
to service access (detector cali-
bration or IQ tool is in progress).
Actions on right-hand section of
X-ray Console are still available.
The diaphragm control is locked
because detector calibration is in
progress.
AWS system cannot communi-
cate with Generator.
A service laptop may be con-
nected in place of the AWS, or
the AWS may be powered OFF
for service requirements.
The X-ray Console and the dia-
phragm control are locked due to
service access (detector calibra-
tion in progress). Actions on
right-hand section of X-ray Con-
sole are still available.
The Image Detection Controller
system cannot communicate with
the Generator.
If communication is not recov-
ered in a few minutes, the system
will be in failure mode (error
E19).
Molybdenum focal track is out of
use.
Rhodium focal track is out of use. Power OFF/ON from the X-ray Console. If the
Page no. 169
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1. Press the prep button again.
2. If the problem occurs again at the next acquisi-
tion, power OFF the whole system and wait for the
end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
Wait until end of service access.
If no service in progress, power OFF/ON from the
X-ray Console.
Wait until end of calibration.
If no calibration in progress, power OFF/ON from
the X-ray Console.
1. Wait for completion of the service operation if
there is one, or wait one minute to allow the system
to recover.
2. If the problem persists, power OFF the whole
system and wait for the end of the shutdown
sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
Wait until end of service access.
If no service in progress power OFF/ON from the
X-ray Console.
Wait until the Image Detection Controller system
recovers.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
You can continue using rhodium focal track.
problem continues, note ALL messages and con-
tact GEMS Field Service.
You can continue using molybdenum focal track.
Error messages
Chapter 16