GE Senographe DS Operator's Manual page 170

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Error messages
Message on
X-ray Console
I23 Console communi-
cation fail
I24 Line power interrup-
tion
I25 Recumbent patient
view names
I27 Generator CPU bat-
tery error
I28 Max wait for same
exposure mn
I29 Excessive mAs for
this mode mAs
I31 AOP abort: use
manual mode
or
I31 Abort: Image signal
too low
Chapter 16
Explanation
Generator-X-ray Console con-
nection is faulty.
Micro power cut occurred on line
supply.
Operator has selected a view
name for recumbent patient.
CPU backup battery is dis-
charged.
It is necessary to wait the dis-
played time to be able to repeat
the same exposure.
The amount of mAs necessary to
repeat the same exposure is too
high in the mode selected.
This error indicates that the aver-
age signal level of the pre-expo-
sure image is too low to calculate
the X-ray parameters for the
exposure.
Page no. 170
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
Power OFF/ON from the X-ray Console.
If problem continues, note ALL messages and con-
tact GEMS Field Service.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
If you do not want to make the acquisition with the
selected view name, enter and leave the Setup/
View menu without pressing Valid, to return to nor-
mal view name selection.
The view name selection will return to normal
selection after the next exposure.
Contact GEMS Field Service.
DO NOT REMOVE POWER FROM EQUIPMENT.
Wait the required amount of time to repeat expo-
sure. If an exposure is attempted before this time,
an abort may possibly occur (there is a safety mar-
gin to ensure proper operation).
You can ignore the message if a different, less
dense, breast is to be imaged next.
Switch to a different mode or to a different filter.
Switch to Manual mode, or ensure that conditions
are correct for use of AOP.
When using AOP:
- Do not use AOP if the breast thickness exceeds
9 cm (3.5 inches).
- Ensure that no highly attenuating objects (mark-
2
ers larger than 2 mm
, needles, etc.) are inside the
AOP sensing zone
- Ensure that the collimator blades are outside the
FOV.
- Ensure that the compression force and thickness
measurements are correctly calibrated.
-Do not use AOP with a breast containing an
implant.
If the problem recurs frequently after taking the
above precautions, note ALL messages and inform
GEMS Field Service.
16-errors.fm