Table 1 Suggested manual techniques
Compressed
Homogeneously fatty
breast
Target/
thickness
Filter
(mm)
With grid (contact examination):
< 30
Mo/Mo
30-40
Mo/Mo
40-50
Rh/Rh
50-60
Rh/Rh
60-70
Rh/Rh
70-80
Rh/Rh
> 80
Rh/Rh
Without grid (magnification examination):
< 30
Mo/Mo
30-40
Mo/Mo
40-50
Mo/Rh
50-60
Rh/Rh
60-70
Rh/Rh
70-80
Rh/Rh
> 80
Rh/Rh
Table 2 Maximum and recommended kV ranges
TRACK
Mo
Mo
Rh
Chapter 12
50% fatty, 50% dense
kVp
mAs
Target/
Filter
25
32
Mo/Mo
26
36
Mo/Rh
28
50
Rh/Rh
29
56
Rh/Rh
29
71
Rh/Rh
29
80
Rh/Rh
30
90
Rh/Rh
25
25
Mo/Mo
26
36
Mo/Mo
28
56
Rh/Rh
29
71
Rh/Rh
30
71
Rh/Rh
31
80
Rh/Rh
32
80
Rh/Rh
FILTER
Mo
Rh
Rh
Page no. 130
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
kVp
mAs
26
28
26
45
29
56
29
63
29
80
30
90
30
140
26
36
27
56
29
71
30
80
31
80
32
90
33
110
KV RANGE
MAXIMUM
RECOMMENDED
22 thru 32
25 thru 28
22 thru 40
26 thru 30
25 thru 49
28 thru 35
Homogeneously dense
Target/
kVp
mAs
Filter
Mo/Mo
26
36
Mo/Rh
27
50
Rh/Rh
29
63
Rh/Rh
30
71
Rh/Rh
30
80
Rh/Rh
31
80
Rh/Rh
31
140
Mo/Mo
27
56
Mo/Rh
27
63
Rh/Rh
30
71
Rh/Rh
31
71
Rh/Rh
32
80
Rh/Rh
33
80
Rh/Rh
35
125