GE Senographe DS Operator's Manual page 185

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Message on
Gantry Readout
A05 Compression
upper position is
reached
A06 Lift end position is
reached
A07 Lift is not opera-
tional
A08 Max manual com-
pression force is
reached
A09 Max power driven
compression force is
reached
A10 Motion stopped.
Multiple movements
are requested
A12 Move tube head to
center position
A14 Risk of collision
with floor
16-errors.fm
Explanation
1) Compression release move-
ment is stopped because the
upper position with the current
paddle is reached.
or
2) The upper position of the pad-
dle holder is reached.
Lift up or down movement is
stopped because the upper or
lower position is reached.
Lift movement is not allowed
because the lift device is not oper-
ational.
The system informs the operator
that the maximum force allowed is
reached.
The system informs the operator
that the maximum force allowed
for motorized movement is
reached.
The current movement is stopped
because another non-compatible
movement is requested:
This advisory message occurs:
A. If you press a user interface
(button or footswitch) during a
pre-set movement.
B. If you press a pre-set button
while a previous movement is in
progress (including deceleration
phase).
Rotation movement has been
requested with the tube head
angled away from the center posi-
tion (0°).
Lift or Rotation movement is
stopped to avoid a collision with
the floor.
Page no. 185
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1) If you have a paddle, remove it to go to the pad-
dle holder upper position.
or
2) If the paddle is removed and the paddle holder
is not at upper position, note ALL messages and
contact GEMS Field Service.
Move the lift in the opposite way.
None
No more compression is available, even using
manual compression.
1) Use the manual control to compress more.
2) If the maximum power driven force is < 20daN
use the Medical Menu to set a higher value.
Release all footswitches or buttons and wait for all
movement to stop before pressing on the foot-
switch or the gantry control movement button
again.
1) If the Stereotactic Positioner is installed, switch
back to angulation mode and move tube head to
the center position (0°).
2) If the Stereotactic Positioner is not installed, an
error has occurred:
A. Power OFF/ON from the X-ray Console.
B. If the problem persists, note ALL messages
and contact GEMS Field Service.
Select another incidence.
Error messages
Chapter 16