Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Browser
Image Process.
Use this screen to enable or disable FineView and/or Premium View.
If FineView is enabled, it will be applied on all RAW images generated by the Medical Application. Refer
to Chapter
12 Image Acquisition
tion on using FineView.
If Premium View is enabled, it will be applied to all eligible processed images generated by the Medical
Application. Refer to Chapter
more information on using Premium View.
If Premium View is disabled, Thickness Equalization will be applied to all eligible processed images
generated by the Medical Application. Refer to Chapter
cessing on page 135
for more information on Thickness Equalization.
Auto Push
Auto Print
Save
8-4
Edit Patient
This option is normally turned off, but may be turned on by a GE Service Engineer if required. It allows
information on the patient Medical Procedure Card (described in Chapter
modified after the exam has been closed. Changes are made by creating a new patient or a new exami-
nation; the original file is always kept for security reasons.
Any changes made to the Patient identity information cause a new patient file to be created, with copies
of the existing exams. Fields which can be changed are: name, ID, date of birth, and sex. No change can
be made to fields such as Operator, Referring Physician, or Performing Physician. When changes are
made and a save is requested, a confirmation message is displayed. If the save is confirmed, a message
is displayed as a warning that a new entry will be made in the database, and suggests that the old infor-
mation should be manually removed. Note that if the old records are not deleted, repeated use of this
facility can rapidly use up disk space.
Chapter 8
Procedure, section
12 Image Acquisition
MedAppPrefs
Annotations
Auto Windowing
Quality Check
FineView
Enable
Disable
Page no. 80
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
9 Image acquisition on page 133
Procedure, section
12 Processing on page 135
12 Image Acquisition
Auto Delete
Test Hosts
Premium View
Enable
Disable
9
for more informa-
Procedure, section
12 Pro-
Names
Image Process.
Close