GE Senographe DS Operator's Manual page 150

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Interchange Media
2-2-2
Preview option
After clicking on the OK button or after using "drag-and-drop" to start recording, but before saving starts,
the Preview of CD-R content option automatically displays a list of the patients which will fit on the CD-R,
and the estimated time required. Note that the list may not be in the order selected by the user, and that
It cannot be assumed that all patients selected for recording will fit.
At this stage, you can still cancel the save operation and modify your selection if required.
To set this option on or off, click the CD-R icon in the Network/Archive panel on the
main Browser and select Option in the drop-down menu.
2-2-3
Checking the CD-R
You are recommended to check for successful completion by browsing the CD-R at the end of the save
operation.
To do so, click the CD-R icon on the main Browser, then select Query in the drop-down menu. If
all the patients selected for the save operation are not displayed on the Media Browser an error
has occurred while recording the CD, and you will have to repeat the save operation.
An error message appears next to the patient name if the associated information has not been
saved. If this occurs:
-
Make a list of the patients correctly saved and those with an error.
-
Insert a new CD-R and repeat the save process.
-
If there are still errors, consult GE Service.
2-2-4
System not available while writing to CD-R
No other operation can be performed on the system while saving images on a CD-R. For a full CD-R the
save operation can take up to 25 minutes.
2-2-5
Errors
If a problem occurs, an error message is displayed:
The following message may be displayed during preparation for saving files; it needs specific action:
Save selection exceeds CD capacity. Save aborted.
Explanation: You have selected more patient files than can be stored on a single CD.
Action: Select fewer patients or fewer exams to be saved on the CD.
If a problem occurs during the save process, the error message Save failed is displayed.
First check that the CD-R drive is switched on, the blank recordable CD is correctly inserted in the
CD-R tray, and the tray is fully closed.
If the items above are correct, an error has occurred while recording the CD, and you will have to
repeat the save operation using a new blank CD.