GE Senographe DS Operator's Manual page 180

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Error messages
Message on AWS screen
BB9
Restart browser must be
performed to take into
account the new configura-
tion
BB10
System currently perform-
ing self-testing - Please
Wait.
CC These messages relate to internal disk system or memory problems:
CC1
ATTENTION Not enough
space to continue
CC2
AWS not available for
acquisition, exam forbid-
den
CC3
Disk space is full, you can-
not perform a new exam
CC4
Image update failed on
disk. You cannot save your
last changes to the exam.
CC5
Medical Preferences Appli-
cation was unable to get
last settings
CC6
Medical Preferences Appli-
cation was unable to save
current settings
CC7
Out of Memory, exam for-
bidden
Chapter 16
Explanation
Medical Application Preferences
settings have been changed,
and Save has been requested.
The changes will not take effect
until the Browser has been
restarted.
The system is busy with internal
checks and tests.
Disk system is full.
The Medical Application cannot
allow acquisition because of
lack of disk system space or an
incomplete action on the AWS.
Disk system is full.
Disk system is full or server has
failed.
The file containing Medical
Application Preferences settings
does not exist or has been cor-
rupted.
The new Medical Application
Preferences settings could not
be saved, probably because of
lack of space.
System requirements exceed
available AWS memory.
Page no. 180
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective action
Restart the Browser.
No action required; wait for completion of
the test.
1. Increase free space on disk system by
deleting exams that have been archived.
2. If the problem persists, contact your
Field Service Representative.
1. Check that all actions are complete
(e.g., close all pop-up windows) and
Increase free space on disk system by
deleting exams that have been archived.
2. If the problem persists, contact your
Field Service Representative.
1. Increase free space on disk system by
deleting exams that have been archived.
2. If the problem persists, contact your
Field Service Representative.
1. Close exam.
2. Increase free space on disk system by
deleting exams that have been archived.
3. If the problem persists, contact your
Field Service Representative.
Note ALL messages and contact your
Field Service Representative.
1. Increase free space on disk system by
deleting exams that have been archived.
2. If the problem persists, contact your
Field Service Representative.
1. Perform system shutdown.
2. Start up system.
3. If the problem persists, contact your
Field Service Representative.