GE Senographe DS Operator's Manual page 194

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Error messages
Message on
Gantry Readout
W27 Positioner power
supply over tempera-
ture
W28 Switching off Gan-
try...
W29 Paddle detection
error
W30 Check sliding pad-
dle position
W31 No defined pad-
dle detected
X01 AWS communica-
tion failure
X02 Bucky communica-
tion failure
X03 Bucky control fail-
ure
Chapter 16
Explanation
Gantry power unit cooling is in a
fault condition; the system risks
being set to an emergency state.
Gantry power OFF sequence is in
progress.
Paddle detection has failed, or
paddle type identification label is
not aligned with the detection win-
dow.
AOP mode (Automatic Optimiza-
tion of Parameters) is inhibited
when the paddle type is not rec-
ognized.
An error occurred during lift move-
ment that does not affect the
safety control but does not allow
knowledge of the absolute lift
position. A consequence of this
error is that rotation movement
will not be stopped in order to
avoid collision with the floor.
The Gantry state cannot allow an
acquisition:
- The Gantry is not communicat-
ing with the AWS.
- It is not possible to perform an
image acquisition.
The Bucky state cannot allow an
acquisition:
- The Bucky is not communicat-
ing.
- It is not possible to perform an
image acquisition.
The Bucky state cannot allow an
acquisition:
- The Bucky device is in a fault
state.
- It is not possible to perform an
image acquisition.
Page no. 194
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1) Note ALL messages and contact GEMS Field
Service.
2) You can continue to use the system but power
OFF from the X-ray Console as soon as you can.
Wait for the end of the power OFF sequence.
1. Remove the paddle and insert it again.
2. If the error persists, try using another paddle.
3. If the problem persists, note ALL messages and
contact GEMS Field Service.
If the paddle is a Senographe DS accessory:
1) Remove the paddle and insert it again.
2) If the message is still displayed after inserting:
A. Power OFF/ON from the X-ray console.
B. If the problem persists, use another
Senographe DS paddle or perform the acquisition
in Manual mode.
C. If the problem persists, note ALL messages
and contact GEMS Field Service.
None. Be careful when performing motion with
tube head near the floor.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, shut down the system
from the Browser.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Unplug the Bucky.
2) Reinstall the Bucky: check that it is correctly
installed and switched on (initialization noise
heard).
3) Power OFF/ON from the X-ray Console.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Unplug the Bucky.
2) Reinstall the Bucky: check that it is correctly
installed and switched on (initialization noise
heard).
3) Power OFF/ON from the X-ray Console.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
16-errors.fm