Error messages
Table 3 Error Messages Displayed on the Readout Panel at the foot of the Gantry
Message on
Gantry Readout
Device removed
Bucky present
1st Spare Bucky
FOV 9x9
FOV 13x18
FOV 13x23
FOV 19x23
1.5 Mag stand present.
Remove face shield if
present!
1.8 Mag stand present.
Remove face shield if
present!
Chapter 16
Explanation
The system informs the operator
that no device is present.
- This message is displayed for 5
seconds when the Bucky or the
Stereotactic Positioner is
removed.
The system informs the operator
that the Bucky is present.
The system informs the operator
that the Bucky detected is not the
standard one. It has detected a
type of Bucky that should not be
used with the current version of
the software.
The system informs the operator
that the selected field of view is
9cm by 9cm.
The system informs the operator
that the selected field of view is 13
cm deep and 18 cm wide.
The system informs the operator
that the selected field of view is 13
cm deep and 23 cm wide.
The system informs the operator
that the selected field of view is 19
cm deep and 23 cm wide.
The system informs the operator
that the magnification platform
(1.5) is present and reminds the
operator that the face shield must
be removed for the diagnostic
application.
The system informs the operator
that the magnification platform
(1.8) is present and reminds the
operator that the face shield must
be removed for the diagnostic
application.
Page no. 182
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
None.
Note that the message is displayed for 5 seconds.
If the Bucky is not installed, an error has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
If the system detects this, an error has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
Press the Field of View button if you want to set
another field of view.
Press the Field of View button if you want to set
another field of view.
Press the Field of View button if you want to set
another field of view.
Press the Field of View button if you want to set
another field of view.
1) If the magnification platform (1.5) is installed,
remove the face shield if not yet done.
2) If the magnification platform is not installed, an
error has occurred:
a) Power OFF/ON from the X-ray Console.
b) If the problem persists, note ALL messages
and contact GEMS Field Service.
1) If the magnification platform (1.8) is installed,
remove the face shield if not yet done.
2) If the magnification platform is not installed, an
error has occurred:
a) Power OFF/ON from the X-ray Console.
b) If the problem persists, note ALL messages
and contact GEMS Field Service.