GE Senographe DS Operator's Manual page 76

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Browser
Annotations
Select the level of screen and print annotation desired. Screen annotations are chosen from Full, Partial,
None, or Custom; the list of annotation types at the middle of the window shows the current default
selections for the selected level. For print annotations, select the model (1, 2, or 3) to be used by default
(refer to Chapter
13
Printing).
Note:
Only annotation Model 3 contains all the information required by MQSA Quality Mammography
Standards.
Auto Push
Auto Print
Save
Auto Windowing
Images are stored with four windowing levels, Standard, High, Low, and User. Select which level is to be
used as the default when an image is displayed. When User is selected, the system automatically opti-
mizes the contrast (Window Width) of the image and uses this value as the default User value if it is less
than the Standard value. However, if Premium View is enabled (see
User value is considered as if Standard value was chosen. The User level can be overwritten if the user
saves the image with modified levels.
Auto Push
Auto Print
Save
Chapter 8
Annotations
Auto Windowing
Screen annotations
Full
Partial
None
Custom
Annotations
Auto Windowing
Auto Windowing
Senographe DS Acquisition System
MedAppPrefs
Quality Check
Auto Delete
Patient information
Acquisition date and time
Hospital information
X Ray parameters
Anatomical information
Processing information
Display parameters
Measurements
Image Process.
MedAppPrefs
Quality Check
Auto Delete
Standard
High
Low
User
Page no. 76
Operator Manual 5307907-3-S-1EN
Test Hosts
Names
Image Process.
Printing Annotations
Model 1
Model 2
Model 3
Close
below), selecting
Test Hosts
Names
Image Process.
Close
8-browser.fm