GE Senographe DS Operator's Manual page 173

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Message on
X-ray Console
S09 No compression
detected
S10 Generator cooling
mn
S100 PLEASE WAIT ...
S11 Tube cooling mn
S12 Configuration not
allowed
S13 Backup mAs < 50
exposure inhibited>
S14 Maximum mAs
exceeded
S20 Acq. Abort, image
can be lost
S21 Waiting for AWS
boot
S22 Waiting for IDC
boot
S23 Laterality is not
selected
16-errors.fm
Explanation
Breast is not compressed.
Generator requires xxxx minutes
to cool.
Generator requires xxxx minutes
to cool.
Tube requires xxxxx minutes to
cool.
Incompatible settings of magnifi-
cation, grid, and focal spot have
been selected.
Maximum mAs < 50.
Exposure inhibited.
Maximum mAs set at installation
are exceeded. This message
appears only if a value below the
maximum possible mAs value
was set at installation.
Exposure has been fully per-
formed but image may be lost
because an error occurred in the
image chain detection or pro-
cessing.
Communication with AWS sys-
tem is not yet established.
Communication with Image
Detection Control system is not
yet established.
You have not selected the lateral-
ity of the breast to perform acqui-
sition. Exposure is inhibited.
Page no. 173
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
Breast compression is mandatory for AOP.
Wait for cooling period (or select lower mAs values
in Manual Mode).
Wait for cooling period (or select lower mAs values
in Manual Mode).
Wait for cooling to terminate (or select lower mAs
values in Manual Mode).
Select compatible settings.
Wait for cooling to terminate (or select lower kV
values).
Reduce the mAs value.
1. Press the Exposure Interrupt button.
2. Check image quality if image is displayed.
3. Start exam again.
4. If the problem persists, power OFF the whole
system and wait for the end of the shutdown
sequence.
5. Restart the system.
6. If the problem occurs again, note ALL messages
and inform GEMS Field Service.
1. Wait until AWS boot and login is finished.
2. If message persists 10 min after the AWS login,
power OFF the whole system and wait for the end
of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
1. Wait until the Image Detection Control boot is
finished.
2. If message persists for more than 10 min, power
OFF the whole system and wait for the end of the
shutdown sequence.
3. Restart the system.
4. If the problem occurs again, note ALL messages
and inform GEMS Field Service.
Select laterality on the X-ray Console. If you are
imaging a test object, the laterality chosen is not
important.
Error messages