GE Senographe DS Operator's Manual page 176

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Error messages
Message on
X-ray Console
S82 Image detection
failure
S83 Detector environ-
ment not OK
S84 Image quality is
degraded
S85 Detector environ-
ment failure
S86 AWS/IDC Commu-
nication lost
S87 AWS/IDC Commu-
nication failure
S88 AWS image deliv-
ery failure
S89 AWS not available
for acquisition
S90 AWS not available,
close exam
Chapter 16
Explanation
One of the image detection com-
ponents has failed.
The detector environment is not
correct for acquisition.
The temperature has drifted from
the optimal value specified for
best image quality.
Detector environment control
has failed.
AWS system cannot communi-
cate with the Image Detection
Controller system but communi-
cation should recover.
AWS to Image Detection Control-
ler system communication has
failed.
AWS image delivery to medical
application failure.
The AWS cannot allow acquisi-
tion because of lack of disk space
or an incomplete action on AWS.
No more AWS resources to con-
tinue.
Page no. 176
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1. Switch OFF the whole system and wait for the
end of the shutdown sequence.
2. Restart the system.
3. If the problem persists, note ALL messages and
inform GEMS Field Service.
1. Wait for completion of system setting.
2. If this X-ray inhibition stays displayed for more
that 10 min., switch OFF the whole system and
wait for the end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
Check image quality on current exam.
If the problem persists, note ALL messages and
inform GEMS Field Service.
1. Switch OFF the whole system and wait for the
end of the shutdown sequence.
2. Restart the system.
3. If the problem persists, note ALL messages and
inform GEMS Field Service.
Wait until the Image Detector Controller system
recovers.
1. Switch OFF the whole system and wait for the
end of the shutdown sequence.
2. Restart the system.
3. If the problem persists, note ALL messages and
inform GEMS Field Service.
1. If this message stays displayed more that 5
min., switch OFF the whole system and wait for the
end of the shutdown sequence.
2. Restart the system.
3. If the problem persists, note ALL messages and
inform GEMS Field Service.
1. Check that all actions are complete.
2. Increase free space on image disk by deleting
exams that have been archived.
3. If the problem persists, note ALL messages and
inform GEMS Field Service.
1. Close exam.
2. Increase free space on image disk by deleting
exams that have been archived.
3. If the problem persists, note ALL messages and
inform GEMS Field Service.