GE Healthcare
Revision 1
Maintenance
3-7
Evaluation
•
For the flexible compression paddle, which is not intended to be flat and parallel to the breast support
table during compression, the deflection must be within the following values:
-
Minimum: 12 mm.
-
Maximum: 14 mm.
3-8
Corrective action
•
If the deflection exceeds the specified limits, the paddle should be replaced.