GE Senographe DS Operator's Manual page 164

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Maintenance
3-7
Evaluation
For the flexible compression paddle, which is not intended to be flat and parallel to the breast support
table during compression, the deflection must be within the following values:
-
Minimum: 12 mm.
-
Maximum: 14 mm.
3-8
Corrective action
If the deflection exceeds the specified limits, the paddle should be replaced.
3-9
References
1. I. Fife, "The physical dimensions of the compressed breast," British Jnl. of Radiology, 64, 73,
1991.
2. Handbook of Glandular Tissue Doses in Mammography, HHS Pub. FDA 85-8239, U. S. Depart-
ment of Health and Human Services, Food and Drug Administration, Rockville, MD, March, 1985.
3. Keith Nelson, Standard Imaging, Middleton, WI, private communication.
Chapter 15
Page no. 164
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
15-maint.fm

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents