GE Senographe DS Operator's Manual page 195

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Message on
Gantry Readout
X04 Collimation com-
munication failure
X05 Collimation control
failure
X06 Generator commu-
nication failure
X07 Invalid bucky type
X09 Compression pad-
dle detection error
X10 X-ray beam form
software error
X11 Tube head com-
munication failure
16-errors.fm
Explanation
The Field Of View cannot be set
at the correct format:
- The Collimator is not communi-
cating.
- It is not possible to perform an
image acquisition.
The Field Of View cannot be set
at the correct format:
- The Collimation function is
failed.
- It is not possible to perform an
image acquisition.
The Gantry state cannot allow an
acquisition:
- The Gantry is not communicat-
ing with the Generator.
- It is not possible to perform an
image acquisition.
The Gantry does not recognize
the Bucky device.
- It is not possible to perform an
image acquisition.
The Gantry does not recognize
the paddle.
- The compression thickness can-
not be computed accurately.
- It is not possible to continue the
current exam, exposure is inhib-
ited.
An error occurred during X-ray
beam format computation by the
Gantry CPU.
- It is not possible to perform an
image acquisition.
The tube focal spot size cannot be
set:
- The tube head device is not
communicating.
- It is not possible to perform an
image acquisition.
Page no. 195
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Unplug the Bucky.
2) Reinstall the Bucky: check that it is correctly
installed and switched on (initialization noise
heard).
3) Power OFF/ON from the X-ray Console.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Remove and reinstall the paddle.
2) If the problem persists, power OFF/ON from the
X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) If you hear the tube tilt motor, the system is try-
ing to recover itself.
2) If the message "S61 X-ray beam not ready" is
present more than 60s on the x-ray control con-
sole, power OFF/ON from the X-ray console.
3) If the problem persists, shut down the system
from the Browser.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
Error messages