GE Senographe DS Operator's Manual page 191

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Message on
Gantry Readout
H30 Please turn off
control console
H31 Positioner power
supply failure
H32 Tube head control
failure
H34 Tube head com-
munication failure
W01 Internal software
error
W02 Internal software
error
W03 Internal software
error
W04 Internal software
error
W05 Internal software
error
16-errors.fm
Explanation
The system cannot shut down the
generator and X-ray console.
- case A. Following an operator
request on the browser.
- case B. Following a generator
error that requested generator
shutdown.
A system shut down should be
requested manually.
The system is set in emergency
mode:
- Gantry power supply failure; the
system is set in a "halt" state.
-It is not possible to perform any
movement or image acquisition.
The system is in emergency
mode:
- Control command from tube
head has failed, the system is set
to "halt" state.
- It is not possible to perform any
movement or image acquisition.
The system is in emergency
mode:
- Tube head control is not commu-
nicating, the system is set to "halt"
state.
- It is not possible to perform any
movement or image acquisition.
An error occurred on the Gantry
CPU that does not affect the cur-
rent system use.
An error occurred on the Gantry
CPU that does not affect the cur-
rent system use.
An error occurred on the Gantry
CPU that does not affect the cur-
rent system use.
An error occurred on the Gantry
CPU that does not affect the cur-
rent system use.
An error occurred on the Gantry
CPU that does not affect the cur-
rent system use.
Page no. 191
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1) Power OFF on the X-ray console (immediately
in case B!).
2) If the error occurred during a normal shut down:
A. Note ALL messages and contact GEMS Field
Service to inform the field engineer that the system
should be shut down itself.
B. You can continue to use the system but you
must power OFF from the X-ray console at the
next occurrence of this error.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Uncompress the patient using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Uncompress the patient using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
Error messages
Chapter 16